FDA product code · Orthopedic · Class 2
PHX: Shoulder Prosthesis, Reverse Configuration
PHX is FDA's product code for "Shoulder Prosthesis, Reverse Configuration": a Class 2 device classified under 21 CFR 888.3660, reviewed by the Orthopedic panel. FDA flags it as an implant.
510(k) review time for PHX
Across 109 510(k) clearances under PHX decided from 2018-01-01 to 2026-10-06, the median was 139 days from FDA receipt to decision. Half took between 72 and 198 days, and 80% were decided within 218 days. These are past clearances, not a forecast for a new submission.
| Fastest | 20 |
|---|---|
| 25th percentile | 72 |
| Median | 139 |
| 75th percentile | 198 |
| 80th percentile | 218 |
| 90th percentile | 267 |
| Slowest | 496 |
| Year | Median days | n |
|---|---|---|
| 2018 | 127 | 14 |
| 2019 | 144 | 9 |
| 2020 | 143 | 13 |
| 2021 | 143 | 21 |
| 2022 | 198 | 9 |
| 2023 | 154 | 9 |
| 2024 | 139 | 9 |
| 2025 | 72 | 9 |
| 2026 (to 10-06, partial) | 133 | 16 |
Source: openFDA 510(k) records, snapshot of 2026-10-06 (openFDA data as of 2026-09-28). Compare 510(k) review times across all device types, see every code side by side in the Clearance Timeline Estimator, or browse the clearances themselves in PredicateScope.
PHX in the public FDA record
| 510(k) clearances, all years | 205 |
|---|---|
| Device recalls | 68 |
| Establishment listings | 1,277 |
| GUDID device identifiers | 7,120 |
Counts of records, not of devices or events. Adverse event reports for PHX are in MAUDEScope, as counts with the query shown: MAUDE has no denominator, so a report count is not a rate and does not show that a device caused an event.
FDA-recognized standards that cite PHX
8 FDA recognized consensus standards cite PHX, 2 of them recognized only in part, among them ASTM F2028-17; ASTM F2028-25a; ISO 5834-3 Second edition 2019-02. Horizontal standards such as biocompatibility, sterilization and software standards can apply without citing the code.
- 11-330 ASTM F2028-17 (partial recognition) — Standard Test Methods for Dynamic Evaluation of Glenoid Loosening or Disassociation
- 11-433 ASTM F2028-25a (partial recognition) — Standard Test Methods for Dynamic Evaluation of Glenoid Loosening
- 8-495 ISO 5834-3 Second edition 2019-02 — Implants for surgery - Ultra-high-molecular-weight polyethylene - Part 3: Accelerated ageing methods
- 8-496 ISO 5834-4 Second edition 2019-02 — Implants for surgery - Ultra-high-molecular-weight polyethylene - Part 4: Oxidation index measurement method
- 8-497 ISO 5834-5 Second edition 2019-02 — Implants for surgery - Ultra-high-molecular-weight polyethylene - Part 5: Morphology assessment method
- 8-501 ISO 5834-1 Fourth edition 2019-02 — Implants for surgery - Ultra-high-molecular-weight polyethylene - Part 1: Powder form
- 8-514 ISO 5834-2 Fifth edition 2019-02 — Implants for surgery - Ultra-high-molecular-weight polyethylene - Part 2: Moulded forms
- 8-572 ASTM F2565-21 — Standard Guide for Extensively Irradiation-Crosslinked Ultra-High Molecular Weight Polyethylene Fabricated Forms for Surgical Implant Applications
Source: FDA Recognized Consensus Standards, snapshot of 2026-09-01. Search them all in the Consensus Standards Lookup.
Related product codes
- KWS Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented
- PKC Prosthesis, Total Anatomic Shoulder, Uncemented Metaphyseal Humeral Stem With No Diaphyseal Incursion, Semi-Constrained
- QHE Shoulder Arthroplasty Implantation System
- HRS Plate, Fixation, Bone
- MAX Intervertebral Fusion Device With Bone Graft, Lumbar
- NKB Thoracolumbosacral Pedicle Screw System