FDA product code · Radiology · Class 2

OWB: Interventional Fluoroscopic X-Ray System

OWB is FDA's product code for "Interventional Fluoroscopic X-Ray System": a Class 2 device classified under 21 CFR 892.1650, reviewed by the Radiology panel.

510(k) review time for OWB

Across 155 510(k) clearances under OWB decided from 2018-01-01 to 2026-10-06, the median was 115 days from FDA receipt to decision. Half took between 34 and 193 days, and 80% were decided within 219 days. These are past clearances, not a forecast for a new submission.

Days from FDA receipt to 510(k) decision, OWB, decisions 2018-01-01 to 2026-10-06 (n = 155)
Fastest9
25th percentile34
Median115
75th percentile193
80th percentile219
90th percentile261
Slowest462
Median days by decision year, OWB (years with enough clearances to compute). The partial final year counts only submissions already decided, so its median runs low.
YearMedian daysn
20185423
20191749
20205316
20216024
20229021
202315212
202412613
202514822
2026 (to 10-06, partial)22115

Source: openFDA 510(k) records, snapshot of 2026-10-06 (openFDA data as of 2026-09-28). Compare 510(k) review times across all device types, see every code side by side in the Clearance Timeline Estimator, or browse the clearances themselves in PredicateScope.

OWB in the public FDA record

Records naming OWB, from openFDA (snapshot of 2026-09-15)
510(k) clearances, all years325
Device recalls624
Establishment listings557
GUDID device identifiers702

Counts of records, not of devices or events. Adverse event reports for OWB are in MAUDEScope, as counts with the query shown: MAUDE has no denominator, so a report count is not a rate and does not show that a device caused an event.

FDA-recognized standards that cite OWB

5 FDA recognized consensus standards cite OWB, 1 of them recognized only in part, among them NEMA XR 27 Amendment 1-2013; IEC 61910-1 Edition 1.0 2014-09; IEC 60601-2-54 Edition 2.0 2022-09. Horizontal standards such as biocompatibility, sterilization and software standards can apply without citing the code.

  • 12-286 NEMA XR 27 Amendment 1-2013 — X-ray equipment for interventional procedures - User Quality Control Mode
  • 12-290 IEC 61910-1 Edition 1.0 2014-09 — Medical electrical equipment - Radiation dose documentation - Part 1: Radiation dose structured reports for radiography and radioscopy
  • 12-348 IEC 60601-2-54 Edition 2.0 2022-09 (partial recognition) — Medical electrical equipment - Part 2-54: Particular requirements for the basic safety and essential performance of X-ray equipment for radiography and radioscopy
  • 12-351 IEC 60601-2-43 Edition 3.0 2022-12 — Medical electrical equipment - Part 2-43: Particular requirements for the safety and essential performance of X-ray equipment for interventional procedures
  • 12-364 IEC 61223-3-8 Edition 1.0 2024-03 — Evaluation and routine testing in medical imaging departments - Part 3-8: Acceptance and constancy tests - Imaging performance of X-ray equipment for radiography and radioscopy

Source: FDA Recognized Consensus Standards, snapshot of 2026-09-01. Search them all in the Consensus Standards Lookup.