FDA product code · Orthopedic · Class 2

JWH: Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer

JWH is FDA's product code for "Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer": a Class 2 device classified under 21 CFR 888.3560, reviewed by the Orthopedic panel. FDA flags it as an implant.

510(k) review time for JWH

Across 187 510(k) clearances under JWH decided from 2018-01-01 to 2026-10-06, the median was 72 days from FDA receipt to decision. Half took between 30 and 123 days, and 80% were decided within 145 days. These are past clearances, not a forecast for a new submission.

Days from FDA receipt to 510(k) decision, JWH, decisions 2018-01-01 to 2026-10-06 (n = 187)
Fastest17
25th percentile30
Median72
75th percentile123
80th percentile145
90th percentile243
Slowest458
Median days by decision year, JWH (years with enough clearances to compute). The partial final year counts only submissions already decided, so its median runs low.
YearMedian daysn
20189025
201913328
20205922
20216124
20225617
20235720
20248919
20255520
2026 (to 10-06, partial)6312

Source: openFDA 510(k) records, snapshot of 2026-10-06 (openFDA data as of 2026-09-28). Compare 510(k) review times across all device types, see every code side by side in the Clearance Timeline Estimator, or browse the clearances themselves in PredicateScope.

JWH in the public FDA record

Records naming JWH, from openFDA (snapshot of 2026-09-15)
510(k) clearances, all years917
Device recalls983
Establishment listings3,462
GUDID device identifiers85,143

Counts of records, not of devices or events. Adverse event reports for JWH are in MAUDEScope, as counts with the query shown: MAUDE has no denominator, so a report count is not a rate and does not show that a device caused an event.

FDA-recognized standards that cite JWH

36 FDA recognized consensus standards cite JWH, among them ASTM F983-86 (Reapproved 2018); ASTM F565-04 (Reapproved 2018); ISO 7207-1 Third edition 2007-02-01. Horizontal standards such as biocompatibility, sterilization and software standards can apply without citing the code.

  • 11-197 ASTM F983-86 (Reapproved 2018) — Standard Practice for Permanent Marking of Orthopaedic Implant Components
  • 11-199 ASTM F565-04 (Reapproved 2018) — Standard Practice for Care and Handling of Orthopedic Implants and Instruments
  • 11-232 ISO 7207-1 Third edition 2007-02-01 — Implants for surgery - Components for partial and total knee joint prostheses - Part 1: Classification definitions and designation of dimensions
  • 11-308 ASTM F3161-16 — Standard Test Method for Finite Element Analysis (FEA) of Metallic Orthopaedic Total Knee Femoral Components under Closing Conditions
  • 11-311 ISO 14243-2 Third edition 2016-09-01 — Implants for surgery - Wear of total knee-joint prostheses - Part 2: Methods of measurement
  • 11-342 ASTM F732-17 (Reapproved 2025) — Standard Test Method for Wear Testing of Polymeric Materials Used in Total Joint Prostheses
  • 11-361 ISO 14243-5 First edition 2019-05 — Implants for surgery - Wear of total knee prostheses - Part 5: Durability performance of the patellofemoral joint
  • 11-364 ISO 14243-1 Second edition 2009-11-15 — Implants for surgery - Wear of total knee-joint prostheses - Part 1: Loading and displacement parameters for wear-testing machines with load control and corresponding environmental conditions for test [Including AMENDMENT1 (2020)]
  • 11-365 ASTM F1800-19 e1 — Standard Practice for Cyclic Fatigue Testing of Metal Tibial Tray Components of Total Knee Joint Replacements
  • 11-366 ASTM F2381-19 — Standard Test Method for Evaluating Trans-Vinylene Yield in Irradiated Ultra-High Molecular Weight Polyethylene Fabricated Forms Intended for Surgical Implants by Infrared Spectroscopy
  • 11-367 ISO 14243-3 Second edition 2014-11-01 — Implants for surgery - Wear of total knee-joint prostheses - Part 3: Loading and displacement parameters for wear-testing machines with displacement control and corresponding environmental conditions for test [Including AMENDMENT1 (2020)]
  • 11-368 ASTM F3334-19 — Standard Practice for Finite Element Analysis (FEA) of Metallic Orthopaedic Total Knee Tibial Components
  • 11-370 ISO 14879-1 Second edition 2020-07 — Implants for surgery - Total knee-joint prostheses - Part 1: Determination of endurance properties of knee tibial trays
  • 11-373 ASTM F1223-20 — Standard Test Method for Determination of Total Knee Replacement Constraint
  • 11-374 ISO 7207-2 Second edition 2011-07-01 — Implants for surgery - Components for partial and total knee joint prostheses - Part 2: Articulating surfaces made of metal ceramic and plastics materials [Including AMENDMENT 1 (2016) and AMENDMENT 2 (2020)]

The first 15 of 36 are shown. Source: FDA Recognized Consensus Standards, snapshot of 2026-09-01. Search them all in the Consensus Standards Lookup.