Guide · Packaging and shelf life

Accelerated aging time: the ASTM F1980 table

One year of shelf life takes 38 days in a 55 °C chamber, two years take 75 days and five years 186 days, using Q10 = 2 and a 22 °C ambient temperature. The tables below cover 6 months to 5 years at 50 to 60 °C, for two common ambient assumptions.

Accelerated aging days, 22 °C ambient

Days in the chamber for each shelf-life claim. Q10 = 2, TRT = 22 °C, rounded up to whole days.
Shelf life50 °C55 °C57 °C60 °C
6 months27191714
1 year53383327
18 months79564940
2 years105756553
3 years1581129779
4 years210149130105
5 years263186162132
Aging factor6.9649.84911.31413.929

Accelerated aging days, 25 °C ambient

Days in the chamber for each shelf-life claim. Q10 = 2, TRT = 25 °C, rounded up to whole days.
Shelf life50 °C55 °C57 °C60 °C
6 months33232017
1 year65464033
18 months97696049
2 years130928065
3 years19413712097
4 years259183159130
5 years323229199162
Aging factor5.65789.1911.314

How the numbers are calculated

ASTM F1980 uses a simplified Arrhenius relation. The accelerated aging factor is AAF = Q10(TAA − TRT) / 10, where TAA is the chamber temperature and TRT the real-time storage temperature. The time in the chamber is the shelf life divided by the AAF. A month here is 30.4375 days (365.25 ÷ 12), and results are rounded up, so a study never comes out short.

Q10 = 2 (the rate of aging doubles for every 10 °C) is the conventional, conservative choice. A higher Q10 shortens the study and needs evidence that the materials support it.

Choosing the chamber temperature

Higher temperatures shorten the study but can change how materials age. A chamber temperature should stay below any temperature at which the packaging or product materials soften, melt or transition. That is why many studies run at 55 °C rather than 60 °C. Humidity is a separate choice: state it in the protocol and justify it for the materials.

Accelerated aging supports a shelf-life claim while real-time aging runs alongside it. The claim is confirmed when real-time samples reach the same age.

Run your own numbers

The Packaging & Accelerated Aging Validator takes any shelf life, chamber temperature, ambient temperature and Q10, builds the pull schedule, and lays out the conditioning (ASTM D4332) and transit (ASTM D4169) tests. The FDA recognitions of ASTM F1980 and ASTM D4169 have their own pages.